Pressure range
| 1. Under normal working conditions, the pressure range of device is 60kPa~130kPa. 2. When the pipeline blockage occurs etc. in the device, the pressure produced by the device will not break in the range of 130kPa ~ 400kPa. |
Machine noise | Noise of machine in normal operation: ≤ 65dB (frequency-weighting characteristic A) |
Continuous working hours | When the device works continuously for more than 4 hours under normal temperature and AC power supply, it should be able to work normally. |
Temperature:5℃~40℃
Relative humidity:≤80%
Atmosphere pressure:80.0kPa~106.0kPa
Spray rate:The spray rate should be ≥0.2mL/min
Air flow volume:Compressed air flow of compression pump:≥8L/min
Residual liquid volume:≤0.7mL
Enclosure leakage current:Normal condition≤0.1mA; Single fault condition≤0.5mA
Patient leakage current:Nomal condition≤0.1mA; Single fault condition≤0.5mA
Droplet partical size:1um-5um
Product size:165*160*95mm
Color box size:208*130*237
Qty in each carton:8
Master carton (WxDxH in mm):670*257*445 mm